Ground resistance meter IC - 10. Operating manual

To protect people from electric shock as a result of touching metal parts of electrical installations, protective grounding is used. This is an artificial connection of equipment housings with a grounding bus. The bus is connected to a grounding loop, which consists of grounding electrodes (grounding electrodes) and grounding conductors. If phase voltage enters the housing, short circuit protection will operate. The machines will de-energize the damaged area, preventing heating and ignition of electrical circuits. Protection saves the life of a person exposed to electric current.


Protective grounding circuit in the ground

2.3.3.1 Resistance measurement using the four-wire method (4P)

Use the “Mode” button to select the four-wire measurement method.

Measuring using the four-wire method excludes from the measurement result the resistance of the measuring cables and the transition resistances at their connection points, which is important when the measured resistance is small. Connect the cables to the measuring sockets T1, P1, P2, T2. Connect to the measured resistance (R) on one side the cables from sockets T1 and P1, and on the other side the cables from sockets P2 and T2 in accordance with Figure 2.3.3.1. Briefly press the “Rx / ¿

».

Software

Current transformers tpl-10-m

The measurement control software is installed in the internal memory of the controller and is not accessible to the user. The metrological characteristics of the device are normalized taking into account the influence of the software.

External software RS-terminal is used to display and present measurement results on an external PC and is not metrologically significant.

Identification data (features) of a metrologically significant part of the software are indicated in Table 1.

Table 1

Software name Software identification name Software version number (identification number) Digital software identifier (executable code checksum) Algorithm for calculating the software identifier
Built-in for IS-20 Firmware 1.0 384E5F73B9A098CE865

B7FFF19C5D658

md5
Built-in for IS-20/1 Firmware 1.0 57992BBA4F719F7FB5D

21B285623B7FA

md5
External software RS-terminal not lower than 1.0

The level of software protection against unintentional and intentional changes is “A”, in accordance with MI 3286-2010.

2.3.3.2 Resistance measurement using the two-wire method (2P)

Use the “Mode” button to select a two-wire measurement method; connect the measuring cables only to sockets T1 and T2. Briefly press the “Rx / ¿

", read the measurement result, consisting of the resistance of the measured object and the resistance of the measuring cables and transition resistances at their connection points. The influence of cable resistance can be eliminated by measuring the resistance of the test cables shorted together, which is then subtracted from the main measurement.

The device allows you to correct the zero of the device - write the resistance of the measuring cables into memory and automatically subtract it from the measurement result. To do this, short-circuit the ends of the measuring cables, enter the “MENU”, select the “CALIBER” function. >0<" and press the "Rx / ¿

" The device will measure the cable resistance and record the result in memory. The measurement result and the inscription “READY” will appear on the indicator. The zero correction operating mode will turn on automatically. In this case, the symbol “>0<” appears on the indicator. When replacing measuring cables, it is necessary to carry out a zero correction again so as not to obtain an incorrect measurement result. To turn off zero correction, enter the “MENU”, select the function “>0< OFF” and press the “Rx /
¿
».

2.3.4 Measuring ground resistance using the four-wire method (4P)

If the controlled object uses its own rules (methods) for measuring grounding resistance, then it is necessary to be guided by them.

Use the “Mode” button to select the four-wire measurement method.

Determine the maximum diagonal (hereinafter D) of the grounding device (GD). Connect the charger using measuring cables to sockets T1 and P1. Install potential pin P2 into the ground at a distance of 1.5 D, but not less than 20 m from the measured charger (Figure 2.3.4).

If there is interference voltage, the device will measure its amplitude value in volts and display the result on the indicator. In this case, it is necessary to find the optimal direction for the location of the measuring pins, at which the magnitude of the interference voltage will be minimal. This will allow you to obtain the most reliable results of subsequent measurements.

Install the current pin T2 into the ground at a distance of more than 3 D, but not less than 40 m from the charger. Connect the connecting cable to the T2 connector of the device. Carry out a series of grounding resistance measurements while sequentially installing the potential pin P2 into the ground at a distance of 10, 20, 30, 40, 50, 60, 70, 80 and 90% of the distance to the current pin T2.

The charger, current and potential measuring pins are usually lined up in one line.

Figure 2.3.4 Ground resistance measurement circuit

four-wire method

Next, plot the dependence of the resistance on the distance between the charger and the potential pin P2. If the curve increases monotonically and has a fairly horizontal section in the middle part (at distances of 40 and 60% the difference in resistance values ​​is less than 10%), then the resistance value at a distance of 50% is taken as the true value.

Otherwise, all distances to the pins must be increased by 1.5–2 times or the direction of installation of the pins must be changed to reduce the influence of overhead or underground communications.

FAQ

What operating conditions are provided for this ohmmeter?
Terms of Use:
  1. For supply voltage, storage and operating temperatures, see the “Technical Specifications” section.
  2. Relative humidity no more than 80% at a temperature of 0...40 °C.
  3. Atmospheric pressure from 630 to 795 mm Hg. Art.
  4. Storage and operation areas should be free of dust, acid, alkali vapors, and gases that cause corrosion.
  5. After being in extreme conditions (storage, transportation), the device is kept in normal (operating) conditions for at least 2 hours.
  6. Power supply: AC voltage (220 ± 20) V, frequency (50 ± 2) Hz
  7. Ventilation openings must not be covered. Minimum distance of 25 mm on the sides.
  8. To clean the outside of the appliance, use a slightly damp cloth. Do not attempt to clean the inside of the appliance. Before cleaning, unplug the device and turn it on only after it has completely dried.
  9. During operation, the following actions are not allowed, leading to refusal of warranty service for the device:
    • Dropping and vibration of the device
    • It is strictly forbidden to apply external voltage to the output connectors of the device.

  10. Do not short-circuit the output or input connectors of the device.
  11. Do not apply external voltage exceeding the permissible voltage to the sync, clock, external modulation, frequency counter, etc. input connectors.
  12. An external signal with the wrong polarity must not be supplied.
  13. It is not allowed to connect a capacitive (without preliminary discharge) or inductive load to the device outputs
  14. When connecting the cable, do not confuse the input connector with the output connector.
  15. This may lead to damage to the device and partial or complete loss of its functionality. Failure of the fuse means a violation of the operating conditions of the device.

Back to section

2.3.5 Measuring ground resistance using the three-wire method (3P)

Using the “Mode” button, select the three-wire measurement method, connect the minimum length measuring cable to the T1 socket. The measurement is carried out similarly to clause 2.3.4, but in this case the resistance of the measuring cable connected to the T1 socket will not be compensated (Figure 2.3.5).

Figure 2.3.5 Ground resistance measurement circuit

three-wire method

2.3.6 Measurement of soil resistivity (Rsp)

The value of soil resistivity is calculated using Werner's measurement method. This technique assumes equal distances between electrodes (d)

, which should be taken at least 5 times the immersion depth of the pins.

Install the measuring pins into the ground in a straight line, at equal distances (d)

and connect to the measuring sockets T1, P1, P2 and T2 in accordance with Figure 2.3.6.
Select the four-wire measurement method mode. Press the “Rx / ¿
” button and read the resistance value
RE
.

The soil resistivity is calculated using the formula: R beat = 2 π • d • RE (6.28 • d • RE).

Figure 2.3.6 — Connection diagram for measuring soil resistivity

To measure with automatic calculation of soil resistivity, use the “Mode” button to select the “R beat” mode, while the previously set distance between the pins is displayed on the indicator.

The distance between the pins can be changed in the “Menu” of the device. Select the function “SET. DIST." The message “DISTANCE BETWEEN ELECTRODES is XXm” appears. Use the ▲ or ▼ buttons to set the distance from 1 to 99 m in 1 m increments. To confirm the selected distance, press the “Rx / ¿

" Measuring soil resistivity is only possible using the four-wire method, which will turn on automatically. The measurement result will be displayed in “mOhm*m”, “Ohm*m” or “kOhm*m”.

The specified distance between the pins is stored in the device’s memory until new values ​​are entered.

2.3.7 Measurements using clamps (if included in the device)

Connect clamps to sockets P1 and P2 (without observing polarity) and select the operating mode with clamps “mA” or “%”.

2.3.7.1 When choosing to measure alternating current “mA”, grasp the circuit under test with clamps. Press and hold the “Rx / ¿

" The device will measure the current and display the result on the screen.

2.3.7.2 Determining the percentage distribution of currents in multi-element grounding allows one to assess the quality of individual grounding devices, including failure (open circuit) of the circuit. A change in this parameter over time makes it possible to assess the nature and rate of aging associated with oxidation of the ground electrode or changes in the condition of the soil.

To determine the percentage distribution of currents through single grounding conductors in multi-element grounding, select this measurement mode “%”. If this operation is carried out without determining the total resistance, then the distance from the charger to the current pin for circuit T2 is selected according to the method of paragraph 2.3.4.

It is recommended to periodically calibrate the clamps before taking measurements. To do this, connect the T1 and T2 sockets to each other, and grab the shorting cable with pliers, select the “Caliber” function in the menu, briefly press the “Rx / ¿

"
The device will calibrate and the “READY” message will appear on the indicator. ¿
button again . However, if there is no contact in the T1T2 circuit or the clamps are not connected (faulty), the device will display a message, respectively, “NO CIRCUIT T1T2” or “NO SIGNAL P1P2”. The message “NO CIRCUIT T1T2” may also appear if the battery is severely discharged.

Connect the measuring leads and clamps according to Figure 2.3.7.2. Grasp a single grounding conductor with pliers, briefly press the “Rx / ¿

" The device will measure the current in the circuit under test, convert it into a percentage distribution and display the result on the screen. If there is no contact in the T1T2 circuit, the device will display the message “NO CIRCUIT T1T2”. Carry out a similar operation with other single grounding conductors.

Figure 2.3.7.2 Determination of percentage current distribution

through single grounding conductors

Attention! The RELIABILITY of measurements using clamps is inversely proportional to the soil resistivity and the number of single grounding conductors in a multi-element grounding.

After working with pliers, for the safety of the device, be sure to

switch to resistance measurement mode or turn off the device.

3 Maintenance and troubleshooting

Maintenance comes down to compliance with the rules of operation and storage.

The device uses a sealed, maintenance-free lead-acid battery with a nominal voltage of 12 V and a capacity of up to 0.8 A/h. Figure 3 shows the polarity of connecting the battery in the compartment.

Figure 3 - Battery connection polarity

The list of possible malfunctions and methods for eliminating them are given in Table 3.

Table 3 - List of possible faults and methods for their elimination

Type of malfunction Probable Cause Elimination method
The device does not turn on Battery is faulty or low Check the condition of the battery, charge or replace it if necessary
The device does not measure resistance Broken connecting cable, oxidation of contacts Check the integrity of the cable, if necessary, repair or replace it, clean the contacts

Repair of the device is permitted only at the manufacturer or specialized repair facilities.

4 Transportation and storage

The device is transported in standard packaging by all types of transport, except for leaky unheated compartments of the aircraft.

Climatic conditions of transportation and storage within the air temperature range from minus 50 to plus 70 ºС with relative air humidity not exceeding 95% at temperatures up to plus 30 ºС and atmospheric pressure from 70 to 106.7 kPa (from 537 to 800 mm Hg .). Exposure to precipitation is not allowed.

Disposal of the device is carried out by the operating organization and is carried out in accordance with the rules and regulations in force in the Russian Federation.

The device does not contain environmentally hazardous elements.

6 Verification

Primary and periodic verifications are carried out by state metrological service bodies or accredited metrological services of legal entities in accordance with PR 50.2.006. Periodic verification is carried out at least once a year, as well as after repairs.

Description of the IS-10 meter

To maintain grounding devices in working condition, the resistance of the circuit located in the ground is measured twice a year. Measurements are carried out during the period of greatest drying of the soil and the greatest freezing of the soil. Grounding is divided into:

  • working;
  • protective;
  • grounding of lightning rods.

Regardless of the type of grounding, they are all connected to electrodes driven into the ground. After measurements are taken, a protocol is drawn up. Areas of grounding devices with measurement results that do not meet the standards must be inspected and repaired.

Attention! The resistance of the local grounding loop for private multi-apartment households should be no higher than 30 Ohms when operating using the TN-CS system (three phases, working zero and grounding conductor).

One of the devices that allows such measurements is the IS-10 grounding resistance meter.


Resistance meter IS-10

The device is used to test resistance values:

  • grounding electrodes (pins);
  • conductors connecting the electrodes to each other and the bus;
  • connection points in the circuit.

With the additional use of a power supply up to 300 V/50 Hz and a current clamp, the amplitude of the alternating current can be found. The device allows you to determine the resistivity of soil and metal compounds.

Table 6.2 — Verification operations

the name of the operation Verification point number
1 External inspection 6.6.1
2 Testing 6.6.2
3 Determination of the limit of the main relative error when measuring resistance 6.6.3
4 Determination of the limit of the main relative error when measuring alternating voltage 6.6.4
5 Determination of the limit of the main relative error when measuring alternating current (if there are clamps in the device) 6.6.5

Note - In the case of a separate supply of KTI-10 clamps, the registration number is entered by the user into the acceptance certificate (clause 7) and an extraordinary verification of the device is carried out according to clause 6.6.5.

6.3 Verification means

When performing verification, the verification tools specified in Table 6.3 must be used.

Table 6.3 — Verification tools

Name and type of verification tool Required technical characteristics of the verification device
Resistance store P4834 from 0.01 to 105

Ohm, up to 50 kHz, CT 0.02

Installation U300 TU77 up to 1000 V
Voltmeter V7-38 ХВ2.710.031TU up to 1000 V, CT 0.4
Device Ts4352 Current up to 6 A, CT 1.0 for DC and 1.5 for AC

Note - Equipment and measuring instruments can be replaced with similar ones that provide the required measurement accuracy of the relevant parameters and measurement limits.

What's included in the package

The is 10 device is equipped according to the list included in the package:

  • IS-10 device – 1 pc.;
  • user manual – 1 pc.;
  • power supply unit BPN-A 12-0.5 – 1 pc.;
  • clamp – 1 pc.;
  • clamps – 2 pcs.;
  • set of connecting cords: 2 pcs. 1.5 m each and 2 pcs. 40 m each (on a reel);
  • storage case – 1 pc.;
  • grounding electrode (L = 1 m) – 4 pcs.;
  • current clamps – 1 pc.

Information. A set of electrodes and current clamps are optional elements of the kit. They are supplied separately. The external power supply may differ from the declared type, but has the necessary parameters.

Rating
( 1 rating, average 4 out of 5 )
Did you like the article? Share with friends:
For any suggestions regarding the site: [email protected]
Для любых предложений по сайту: [email protected]